4-6
WARNING
z
When the filling level of the anesthetic agent is outside the maximum or minimum
mark, incorrect output concentration may occur.
z
Before use, check the filling level of the anesthetic agent.
CAUTION
z
Anesthetic agent vapor will escape into ambient atmosphere if filling operations are not
done properly.
4.2.2 Key Filler System
The filling steps of V60 Isoflurane Key Filler Vaporizer and V60 Sevoflurane Pour Fill Vaporizer are
the same.
If the vaporizer is connected to the anesthetic delivery system, fresh gas flow can remain as set.
1.
Turn the control dial clockwise back to “0” position until the “0” button pops up.
WARNING
z
Significant quantities of anesthetic agent vapor may escape if the control dial does not
return to “0” position.
Turn the control dial back
to “0” position