10-1
10
Storage and Transport
10.1 Storage
Storage for longer than 6 months:
1. Drain and blow off the vaporizer (See
4.3 Draining the Vaporizer
and
4.4 Blowing off the
Vaporizer
).
2. Turn the control dial to “0”. The vaporizer handle for locking lever and interlock device are
in their original positions.
3. If packing is necessary, see
10.2 Transport
.
4. Observe storage temperature. See
A Product Specifications
. If storage temperature range is
exceeded, internal damage may occur which could cause incorrect output concentration.
Before putting into operation again, carry out all-round inspection first.
NOTE
z
When the anesthetic vaporizer is not in use for a long period of time, use the plug to
block the gas inlet and outlet.
z
If the anesthetic vaporizer is stored in high temperature and then used, the
concentration of the delivered anesthetic agent may be high. To enable pressure
equalization, always turn the control dial to 1% after connecting the anesthetic
workstation, and wait for at least 15 seconds.
Summary of Contents for V60
Page 1: ...V60 Anesthetic Vaporizer Operator s Manual ...
Page 2: ......
Page 9: ...3 C Symbols and Terminology C 1 C 1 Symbols C 1 C 2 Terminology C 1 ...
Page 10: ...4 FOR YOUR NOTES ...
Page 17: ...2 3 2 3 2 Rear View 1 Handle for locking lever 2 Interlock system 3 Locking pin ...
Page 48: ...5 4 FOR YOUR NOTES ...
Page 56: ...6 8 FOR YOUR NOTES ...
Page 66: ...9 4 FOR YOUR NOTES ...
Page 79: ...11 11 0 1 2 3 4 5 6 0 1 2 3 4 5 6 G C B A F Concentration delivered vol Enflurane Time h ...
Page 80: ...11 12 ...
Page 84: ...A 4 FOR YOUR NOTES ...
Page 86: ...B 2 FOR YOUR NOTES ...
Page 89: ......
Page 90: ...P N 046 001002 00 8 0 ...