4-16
4.3.2 Key Filler System
The draining steps of V60 Isoflurane Key Filler Vaporizer and V60 Sevoflurane Pour Fill Vaporizer are
the same.
Place the vaporizer upright or suspend so that all the anesthetic agent can drain out.
1. Turn the control dial clockwise back to “0” position until the “0” button pops up.
2. Select the correct anesthetic agent bottle and open the bottle. Do not use a damaged filling
adapter or anesthetic agent bottle.
3. Select the correct filling adapter for the anesthetic agent.
4. Screw the filling adapter firmly into the anesthetic agent bottle.
5. Turn the lock knob counter clockwise.
CAUTION
z
If the connection between the filling adapter and anesthetic agent bottle is not
leak-tight, anesthetic agent may escape, endangering health.