9-2
vaporizer is defective.
The vaporizer is incorrectly
installed or the plug-in adapter is
damaged.
If necessary, re-install the vaporizer
or have it repaired by trained service
personnel.
The vaporizer detection
system on anesthetic
delivery system displays
anesthetic agent which is
different from the
vaporizer.
A different anesthetic agent has
just been used and high
concentrations of it are still
present in the breathing system.
Flush the breathing system or wait
for gas to change.
The monitor settings have not
been changed after anesthetic
agent has been changed.
Change monitor settings.
The control dial cannot be
set to concentration.
Interlock jams or another
vaporizer is still switched on.
Switch off other vaporizer.
For interlock fault, have it repaired
by trained service personnel.
The “0” button is not pressed.
Press the “0” button.
The control dial is jammed.
Repair by trained service personnel.
The concentration can be
adjusted without pressing
the “0” button.
The “0” button is defective.
Repair by trained service personnel.
Anesthetic agent vapor
has leaked during use.
The plug-in adapter is not fitted
flush.
Check the anesthesia machine
plug-in connector sealing rings and
sealing surfaces. Check that the
locking lever is not buckled.
The filler cap is not tightened or
the sealing ring is defective.
Tighten the filler cap. Repair by
trained service personnel if the
sealing ring is defective.
Drainage screw is not closed.
Tighten the drainage screw.
Filling level cannot be
read in the sight glass or
incorrect filling level is
shown in the sight glass.
The vaporizer is completely
empty.
Refill the vaporizer.
The vaporizer is overfilled.
Drain the vaporizer to the maximum
mark and check the concentration.
Sight glass display is faulty.
Repair by trained service personnel.
Summary of Contents for V60
Page 1: ...V60 Anesthetic Vaporizer Operator s Manual ...
Page 2: ......
Page 9: ...3 C Symbols and Terminology C 1 C 1 Symbols C 1 C 2 Terminology C 1 ...
Page 10: ...4 FOR YOUR NOTES ...
Page 17: ...2 3 2 3 2 Rear View 1 Handle for locking lever 2 Interlock system 3 Locking pin ...
Page 48: ...5 4 FOR YOUR NOTES ...
Page 56: ...6 8 FOR YOUR NOTES ...
Page 66: ...9 4 FOR YOUR NOTES ...
Page 79: ...11 11 0 1 2 3 4 5 6 0 1 2 3 4 5 6 G C B A F Concentration delivered vol Enflurane Time h ...
Page 80: ...11 12 ...
Page 84: ...A 4 FOR YOUR NOTES ...
Page 86: ...B 2 FOR YOUR NOTES ...
Page 89: ......
Page 90: ...P N 046 001002 00 8 0 ...