8-4
5. After
test
(1) Switch off the vaporizer. Set the control dial to “0” position.
(2) Switch off Air or O
2
flow on the anesthesia machine.
8.4 Checking the Filling System
Verify the following
:
Pour Fill system
1. The sealing ring for filler cap is in good condition.
2. The filling opening is clean.
3. The sight glass shows normal liquid level.
Key Filler system
1. The sealing cushion for filling device is in good condition.
2. Only the correct filling adapter fits into the filling system.
3. The filler control can be opened and closed smoothly.
4. The sight glass shows normal liquid level.
Quik-Fil system
1. The sealing ring for filler cap is in good condition.
2. The filling opening is clean.
3. The valve core inside the filling opening can be depressed and retracts smoothly.
4. The sight glass shows normal liquid level.
8.5 Checking the Plug-in Adapter
Verify the following
:
1. When the Handle for locking lever is turned to locking position, turns back automatically.
2. The locking lever is undamaged and not buckled.
3. The interlock device is undamaged, guides easily and cannot be removed.
4. Two interlock pins are present.
5. Sealing areas are undamaged.
6. Manufacturer’s plate on the back of the vaporizer is present and secure.
Summary of Contents for V60
Page 1: ...V60 Anesthetic Vaporizer Operator s Manual ...
Page 2: ......
Page 9: ...3 C Symbols and Terminology C 1 C 1 Symbols C 1 C 2 Terminology C 1 ...
Page 10: ...4 FOR YOUR NOTES ...
Page 17: ...2 3 2 3 2 Rear View 1 Handle for locking lever 2 Interlock system 3 Locking pin ...
Page 48: ...5 4 FOR YOUR NOTES ...
Page 56: ...6 8 FOR YOUR NOTES ...
Page 66: ...9 4 FOR YOUR NOTES ...
Page 79: ...11 11 0 1 2 3 4 5 6 0 1 2 3 4 5 6 G C B A F Concentration delivered vol Enflurane Time h ...
Page 80: ...11 12 ...
Page 84: ...A 4 FOR YOUR NOTES ...
Page 86: ...B 2 FOR YOUR NOTES ...
Page 89: ......
Page 90: ...P N 046 001002 00 8 0 ...