SpO
2
Measurements
Operation
3 - 28
Accutorr V Operating Instructions
necessary, remove and reapply the sensor. If any of the above mentioned indications
occur, immediately remove the sensor and find an alternate site.
NOTE:
Check the sensor site more frequently on neonate and active
patients.
• Incorrect placement can also reduce the acquired sensor signal, and therefore
compromise performance. Select an alternate site (toe) if the sensor can not be placed
on the patient’s finger correctly or if the fingernails interfere with the acquisition of a
reliable signal.
• Use of the reusable sensor is not recommended for long-term monitoring (4 to 6 hours).
For monitoring situations exceeding 4 to 6 hours, either reposition the reusable sensor
every 2 to 4 hours to a different site (finger/toe) or use a disposable sensor with its
appropriate bandage.
CAUTION:
• Do not over-tighten the sensor bandages. Excessive pressure on the monitoring site can
affect SpO
2
readings and may reduce readings below true SpO
2
. Excessive pressure
can also result in pressure necrosis and other skin damage.
3.11.2
Sequence for Establishing SpO
2
with Nellcor
®
Pulse Oximetry
This is an optional feature.
1.
Plug the sensor directly into the SpO2 Connector (20) or if necessary, use a Nellcor
®
DOC-10 extension cable.
2.
See package insert(s) for use and care instructions. Additional information is available
from Nellcor Puritan Bennett Inc. at www.nellcor.com.