Accutorr V Operating Instructions
6 - 15
Appendix
Electromagnetic Compatibility
6.7
Electromagnetic Compatibility
Electromagnetic Compatibility
NOTE:
The Accutorr V needs special precautions regarding EMC and
needs to be installed and put into service according to the
EMC information provided in the following tables.
NOTE:
Portable and mobile RF communications equipment can
affect the Accutorr V. See TABLE 6-1 through TABLE 6-4.
TABLE 6-1
GUIDANCE AND DECLARATION - ELECTROMAGNETIC EMISSIONS
The Accutorr V is intended for use in the electromagnetic environment specified in the table. The
customer or the user of the Accutorr V should assure that it is used in such an environment.
EMISSIONS
TEST
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
RF emissions
CISPR 11
Group 1
The Accutorr V uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class A
The Accutorr V is suitable for use in all establishments
other than domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes
Harmonic
emissions IEC
61000-3-2
Class A
Voltage
fluctuations/
Flicker
emissions IEC
61000-3-3
Complies