Accutorr V Operating Instructions
3 - 55
Operation
Status and Error Codes
3.17.10
System Alarm Messages
3.17.11
Prompt Messages
ALARM MESSAGE
A
B
LEVEL
CAUSE
ACTION
REAL CLOCK NEED SET
No
No
L
The system time is
incorrect.
Reset the system time and
then restart the monitor.
REAL CLK NOT EXIST
No
No
L
There is no button
battery, or the battery
power is depleted.
Install a button battery or
replace with a new one.
NET INIT ERR (G.)
No
No
L
The monitor cannot be
connected to the
network due to network
problem.
Contact service personnel
for repair.
NET INIT ERR (RAM)
No
No
L
NET INIT ERR (REG)
No
No
L
NET INIT ERR (MII)
No
No
L
NET INIT ERR (LOOP)
No
No
L
NET ERR (RUN 1)
No
No
L
NET ERR (RUN 2)
No
No
L
NET ERR (RUN 3)
No
No
L
12V TOO HIGH
No
No
L
A problem occurs to the
system’s power supply.
If this alarm message
occurs frequently, contact
service personnel for
repair.
12V TOO LOW
No
No
L
BATTERY TOO LOW
No
No
H
The battery voltage is
too low.
Connect the monitor to AC
mains to charge the
battery.
NET ERROR
Yes
No
L
The monitor is not
connected to the
network.
Check the network links.
The “A” field indicates whether all alarm indications can be cleared or not, and the “B” field indicates whether all alarm
indications except the alarm message can be cleared or not. The alarm levels are H = high and L = low.
PROMPT MESSAGE
CAUSE
ACTION
PULSE SEARCH
The SpO
2
module is searching the pulse.
Wait till the end of the search.
REC INITIALIZING
The recorder is initializing
Wait until initialization complete before
printing.
NIBP MEASURING...
NIBP measurement in process
Wait until measurement complete or
press
measurement and deflate the cuff.
PRESS START NIBP
Interval NIBP measurement has been
stopped by pressing
Press
(24) to restart NIBP
interval mode measurements.
PNEUM TESTING...
Pneumatic testing in process.
Wait until testing complete.
MEASUREMENT
COMPLETE
A measurement has finished.
None
NIBP RETRY
NIBP measurement failed and is retrying.
Wait until measurement complete or
press
measurement and deflate the cuff.
NIBP RETRY PUMP HIGHER
NIBP measurement failed and is retrying
with a higher pressure.
BARCODE FAILED
The barcode reader failed to read the
patient’s barcode.
Use another barcode reader. If failure
continues, contact service personnel.