Accutorr V Operating Instructions
6 - 5
Appendix
Specifications
6.2.7
Temperature
Monitor
Predictive
6.2.8
Nellcor
®
Performance Specifications
SpO
2
Range:
77 – 111.2°F, 25 – 44°C
Display Resolution:
0.1°F, 0.1°C
Accuracy:
Meets ASTM E1112-86 for accuracy
Range:
95 – 109.4°F, 35 – 43°C
Display Resolution:
0.1°F, 0.1°C
Range:
70 – 100% SpO
2
Display Resolution:
1% SpO
2
Averaging Time
4–6 seconds
Typical response Time
15 seconds
Alarm Condition delay plus
alarm generation
<1 second
Calibration:
Factory Calibrated to Functional Saturation
Accuracy - Nellcor
®
:
1
Sensors: MAX-A, MAX-AL, MAX-N, MAX-P,
MAX-I, & MAX-FAST: ± 2% from 70 to 100%
Sensors: OxiCliq A, OxiCliq N, OxiCliq P,
OxiCliq I: ± 2.5% from 70 – 100%
Sensors: D-YS, DS-100A, OXI-A/N, OXI-P/I:
± 3% from 70 – 100%
Sensors: MAX-R, D-YSE, D-YSPD: ± 3% from
70 – 100%
<70% unspecified
Pulse Rate
1
When sensors are used on neonatal subjects as recommended, the specified accuracy range
is increased by
±
1 digit, to account for the theoretical effect on oximeter measurements of
fetal hemoglobin in neonatal blood.