Accutorr V Operating Instructions
3 - 53
Operation
Status and Error Codes
3.17.8
SmarTemp
™
TEMP Module Alarm Messages
ALARM MESSAGE
A
B
LEVEL
CAUSE
ACTION
WARMUP TIMED OUT
Yes
Yes
L
TEMP probe initial
temperature is too high.
Cool the TEMP probe
before taking
measurement.
WARMING RESISTOR
ERR
No
No
L
The warming resistor in the
TEMP probe fails.
Replace the TEMP
probe.
TEMP PROBE
MISPLACED
Yes
No
L
TEMP probe is not placed in
the sheath or the probe
sheath is not in place.
1. Check that the probe
sheath is in place.
2. Replace the TEMP
probe in the sheath
properly.
ENV TEMP
OVERRANGE
No
Yes
L
The ambient temperature is
beyond the measuring
range.
Take measurement in
proper working
condition.
TEMP VOLTAGE ERR
No
Yes
L
Supply voltage is too high or
too low.
Check the power
supply.
TEMP NO PROBE
No
Yes
L
The TEMP probe is
disconnected from the TEMP
module.
Reconnect the probe
with the TEMP module.
TEMP PREDICTION
ERR
Yes
Yes
L
Improper temperature
measurement
Take TEMP
measurement again
correctly.
TEMP SELFTEST ERR
No
No
L
An error occurs during the
TEMP module initialization
Replace the TEMP
module.
TEMP PROBE OFF
No
Yes
L
TEMP probe does not
contact with the patient.
Take measurement
again after the probe
warms up.
TEMP OVER HIGH
LIMIT
No
No
L
The temperature measured is
too high or measurement
error
Lower the measured
temperature or replace
the TEMP module.
TEMP OVER LOW
LIMIT
No
No
L
The temperature measured is
too low or measurement
error
Raise the measured
temperature or replace
the TEMP module.
TEMP WRONG PROBE
No
No
L
A TEMP probe not supplied
by Mindray DS USA, Inc./
Shenzhen Mindray Bio-
Medical Electronics Co., Ltd
is used.
Replace with a TEMP
probe Mindray DS
USA, Inc./Shenzhen
Mindray Bio-Medical
Electronics Co., Ltd
supplies.
TEMP COMM ERR
No
No
L
TEMP module is not
available or TEMP module
fails
Check if a TEMP
module is available. If
yes, replace the TEMP
module.
The “A” field indicates whether all alarm indications can be cleared or not, and the “B” field indicates whether all alarm
indications except the alarm message can be cleared or not. The alarm levels are H = high and L = low.