Alarms
Operation
3 - 18
Accutorr V Operating Instructions
Alarm Limit Table
PARAMETER
RANGE
UNITS
FACTORY
DEFAULT
UNITS OF
INCREMENT
Systolic High
Adult
Pediatric
Neonate
Off, 60–235
Off, 60–160
Off, 50–120
mmHg
Off
Off
Off
5
Systolic Low
Adult
Pediatric
Neonate
Off, 55–230
Off, 55–155
Off, 45–115
mmHg
Off
Off
Off
5
Diastolic High
Adult
Pediatric
Neonate
Off, 35–200
Off, 35–150
Off, 25–100
mmHg
Off
Off
Off
5
Diastolic Low
Adult
Pediatric
Neonate
Off, 30–195
Off, 30–145
Off, 20–95
mmHg
Off
Off
Off
5
MAP High
Adult
Pediatric
Neonate
Off, 35–235
Off, 30–160
Off, 25–120
mmHg
Off
Off
Off
5
MAP Low
Adult
Pediatric
Neonate
Off, 30–230
Off, 30–155
Off, 20–115
mmHg
Off
Off
Off
5
NIBP Pulse Rate
High
Adult
Pediatric
Neonate
Off, 37–245
Off, 37–245
Off, 72–245
bpm
Off
Off
Off
1
NIBP Pulse Rate
Low
Adult
Pediatric
Neonate
Off, 35–243
Off, 35–243
Off, 70–243
bpm
Off
Off
Off
1
DPM SpO
2
High
Adult
Pediatric
Neonate
Off, 51–100
Off, 51–100
Off, 51–100
%SpO
2
Off
Off
Off
1
DPM SpO
2
Low
Adult
Pediatric
Neonate
50–99
50–99
50–99
%SpO
2
85
90
92
1
DPM SpO
2
Pulse Rate High
Adult
Pediatric
Neonate
Off, 2–254
bpm
Off
1
DPM SpO
2
Pulse Rate Low
Adult
Pediatric
Neonate
Off, 0–252
bpm
Off
1