Accutorr V Operating Instructions
3 - 49
Operation
Status and Error Codes
NIBP WEAK SIGNAL
No
No
L
Failures occur in the pulse
measurement. The
monitor cannot perform
measurement, analysis,
or calculation.
Check the patient’s
condition and verify
patient type. Replace with
an appropriate cuff and
connect it correctly. If the
problem still exists, contact
service personnel for
repair.
NIBP OVERRANGE
No
No
L
Failures occurred in the
pulse measurement. The
monitor cannot perform
measurement, analysis,
or calculation.
EXCESSIVE MOTION
No
No
L
Excessive motion of the
patient’s arms
OVER PRESSURE
No
No
L
The airway might be
blocked.
SIGNAL SATURATED
No
No
L
A failure occurred during
pulse measurement. The
monitor cannot perform
measurement, analysis,
or calculation.
PNEUMATIC LEAK
No
No
L
Leak in the airway.
NIBP SYSTEM FAILURE
No
No
L
Failures occur in the pulse
measurement. The
monitor cannot perform
measurement, analysis,
or calculation.
NIBP TIME OUT
No
No
L
CUFF TYPE ERR
No
No
L
The cuff used does not
correspond to selected
patient type.
NIBP RESET ERROR
No
No
L
Illegal reset during NIBP
measurement.
Check the airway and
take measurements again.
If the problem still exists,
contact service personnel
for repair.
ALARM MESSAGE
A
B
LEVEL
CAUSE
ACTION
The “A” field indicates whether all alarm indications can be cleared or not, and the “B” field indicates whether all alarm
indications except the alarm message can be cleared or not. The alarm levels are H = high and L = low.