Introduction
Indications For Use
x
Accutorr V Operating Instructions
Indications For Use
The Accutorr V is intended for intra-hospital use under the direct supervision of a licensed
healthcare practitioner. The Indications for Use for the Accutorr V include the monitoring of
the following human physiological parameters:
• Noninvasive blood pressure (NIBP)
• Pulse oximetry (SpO
2
)
• Heart Rate
• Temperature
Product Limitations
Non-invasive blood pressure (NIBP) accuracy depends on the application of the proper cuff
size. See Chapter 3.0 for detailed information.
The Accutorr V will not operate effectively on patients who are experiencing convulsions or
tremors.
The Accutorr V is a portable device intended for intra-hospital use.
If the pressure cuff is not placed at the patient’s heart level, the NIBP measurement may be
subject to error, due to the hydrostatic effect.
The pulse rate data displayed on the Accutorr V is computed from the measurement of
peripheral pulses (peripheral pulses taken only during a measurement cycle). The rate
measured by the Accutorr V may differ from the rate of an ECG monitor. This is because the
ECG is an electrical signal that may not always result in a peripheral pulse.
Administration of certain vasoconstrictor drugs (for example, norepinephrine), may reduce
peripheral perfusion to a level that prevents the Accutorr V from taking pulse rate
measurements.
Arterial compression, tricuspid regurgitation, or other conditions may reduce perfusion to a
level that prevents the Accutorr V from taking pulse rate measurements.
The presence of arrhythmias may increase the time required to complete a measurement and
may extend this time so that a measurement cannot complete.
Mode of Operation
Continuous
Protection Against Hazard of Explosion
Not Protected (Ordinary)
Protection Against Ingress of Liquids
IPX1
Degree of Electrical Connection Between
Equipment and Patient
Equipment designed for direct electrical and
non-electrical connection to the patient.
Degree of Mobility
Portable