Accutorr V Operating Instructions
0070-10-0699-02
4 - 3
User Maintenance
Cleaning and Disinfection of the Accutorr V Monitor
4.2
Cleaning and Disinfection of the Accutorr V Monitor
WARNING: Be sure to shut down the monitor and disconnect all power
cords from the outlet before cleaning.
The equipment should be cleaned regularly. Please consult your hospital’s policy for the
recommended frequency for cleaning and disinfecting equipment.
The exterior surfaces of the equipment may be cleaned with a clean and soft cloth, sponge or
cotton ball, dampened with either of the following cleaning solutions:
• Mild soap (Diluted)
• Quaternary ammonia (Diluted)
• Sodium hypochlorite bleach (10%)
• Hydrogen peroxide (3%)
• Ethyl alcohal (70%)
• Isopropyl alcohol (70%)
• Super sani-cloth (0.5% quaternary a 55% Isopropyl alcohol)
To avoid damage to the equipment:
• ALWAYS use solutions in accordance with the manufacturer's instructions.
• ALWAYS wipe off the excess cleaning solution with a dry cloth after cleaning.
• NEVER submerge the equipment into water or any cleaning solution, or pour or spray
water or any cleaning solution on the equipment.
• NEVER permit fluids run into the casing, switches, connectors, or any ventilation
openings in the equipment.