Accutorr V Operating Instructions
3 - 11
Operation
Manual NIBP Measurements
3.6
Manual NIBP Measurements
The Accutorr V calculates NIBP values using the oscillometric method of noninvasive blood
pressure measurement. These measurements correspond to comparisons with auscultatory
values, measured using the fifth Korotkoff sound within ANSI/AAMI SP10 standards for
accuracy.
1.
Select a pressure cuff that is appropriate for the patient size. Use the following chart as
a guideline.
NOTE:
The Accutorr V cuffs have special quick connect connectors.
WARNING: To ensure proper performance and safety and to prevent
The limb pressure may not fall to zero between measurements if the cuff is wrapped too
tightly. Therefore, assure that the cuff is properly applied.
LIMB
CIRCUMFERENCE*
DESCRIPTION /
CUFF NAME
PART NUMBER
REUSABLE CUFFS – QUICK CONNECT– LATEX FREE*
10 – 19 cm
Small Child
0683-15-0001-01
18 – 26 cm
Small Adult
0683-15-0002-01
25 – 35 cm
Adult
0683-15-0003-01
33 – 47 cm
Large Adult
0683-15-0004-01
46 – 66 cm
Adult Thigh
0683-15-0005-01
25 – 35 cm
Adult Long
0683-15-0006-01
33 – 47 cm
Large Adult Long
0683-15-0007-01
SINGLE PATIENT USE CUFFS – QUICK CONNECT– LATEX FREE**
10 – 19 cm
Small Child
0683-14-0001-01 (box of 10)
18 – 26 cm
Small Adult
0683-14-0002-01 (box of 10)
25 – 35 cm
Adult
0683-14-0003-01 (box of 10)
33 – 47 cm
Large Adult
0683-14-0004-01 (box of 10)
46 – 66 cm
Adult Thigh
0683-14-0005-01 (box of 5)
25 – 35 cm
Adult Long
0683-14-0006-01 (box of 10)
33 – 47 cm
Large Adult Long
0683-14-0007-01 (box of 10)
SINGLE PATIENT USE CUFFS – QUICK CONNECT– LATEX FREE**
(REQUIRES HOSE P/N 0683-04-0003)
3 – 6 cm
Neonatal, Size 1
0683-23-0001 (box of 10)
5 – 8 cm
Neonatal, Size 2
0683-23-0002 (box of 10)
7 – 10 cm
Neonatal, Size 3
0683-23-0003 (box of 10)
9 – 13 cm
Neonatal, Size 4
0683-23-0004 (box of 10)
12 – 17 cm
Neonatal, Size 5
0683-23-0005 (box of 10)
*
The limb circumferences of cuffs adhere to the American Heart Association (AHA) guidelines for size.
They also incorporate index and range lines to assist in cuff selection.
**
Do not reuse single patient use cuffs.