Enter the appropriate data for the filter being entered.
12.
Use Table 5 as a guide.
Table 5: Filter Selection Table
Entry
Optional or
Mandatory
Instructions
Notes
Filter Size
Designation
Mandatory Enter
an
integer from
1–4
If the filter is a 10-inch filter, enter 1; if it is a
20-inch filter, enter 2; if it is a 30-inch filter,
enter 3; and if it is a 40-inch filter, enter 4.
For all other filters, enter 1 (for example, a
5-inch cartridge, a 4-inch capsule, a TFF
cassette would all require 1 to be entered).
Filter Size
Displayed
Optional
Enter the filter
size.
This is a text field that is for descriptive
purposes.
Default
Number of
Cartridges
Mandatory Enter
‘1’
This entry should always be 1. If you plan to
test a system with multiple filters then that will
be specified in the Test editor when the filter
test is programmed.
Pore Size
Optional
Enter the pore
size or
efficiency
rating of the
filter.
This is a text field that is for descriptive
purposes.
Summary of Contents for Integritest Exacta
Page 2: ...Topics in this Manual...
Page 7: ......
Page 8: ...Using Help for the Integritest Exacta Instrument...
Page 11: ......
Page 12: ...Getting Started...
Page 20: ...Hardware Connections...
Page 32: ...First Log On...
Page 48: ...Starting Stopping the Instrument...
Page 58: ...Security Setup...
Page 69: ......
Page 70: ...Managing User Access...
Page 78: ...Instrument Setup...
Page 90: ...Filter Editor...
Page 101: ......
Page 102: ...Integritest Exacta Instrument Tests...
Page 109: ......
Page 110: ...Test Editor...
Page 122: ...Running a Test...
Page 131: ......
Page 132: ...Printing...
Page 139: ......
Page 140: ...Printer Maintenance and Troubleshooting...
Page 147: ......
Page 148: ...Data Management...
Page 171: ......
Page 172: ...Networking...
Page 175: ......
Page 176: ...Maintenance and Calibration...
Page 199: ......
Page 200: ...Troubleshooting...
Page 214: ...Error Messages...
Page 225: ......
Page 226: ...Instrument Repair...
Page 232: ...Warranty...
Page 235: ......
Page 236: ...Integritest Exacta Specifications...
Page 246: ...Interface Conventions...
Page 253: ......
Page 254: ...Contacting Millipore...
Page 262: ...References...
Page 263: ......
Page 265: ......
Page 266: ...Glossary...
Page 271: ......
Page 272: ...Index...