Integritest
Exacta Filter Integrity Test Instrument
Operating and Maintenance Manual
Maintenance and Calibration
183
R
2
Value
Decision
Action
>1.82 x variance
A
t
test on residuals is
conducted on the
calibration coefficients
If the t test fails, the
calibration coefficients (for
slope & the zero intercept)
are automatically adjusted.
If the t test passes, the
transducer(s) need
replacement or factory
adjustment. The unit is
locked out from performing
tests.
Recommended Reference Pressure Gauge
The reference gauge for External Pressure Calibrations must have the following characteristics:
•
NIST traceable (or equivalent)
•
0.02% accurate, Full Scale
•
0–100 psig
Note:
You can order a reference gauge direct from Millipore, catalogue no. XITXSP114.
Calibration Setup
1.
Connect the Integritest Exact inlet tubing to the instrument and to a regulated
compressed gas (nitrogen or air) source.
2.
Set the inlet gas pressure to 100 psig.
3.
Remove the disk test adapter, water trap and instrument valve.
4.
Connect the reference pressure gauge to the Integritest Exacta Outlet port.
5.
Make sure that the reference gauge is properly zeroed before continuing with
procedure.
Note:
Use of long tubing sets or multiple fittings may degrade the accuracy of the
referee gauge. Millipore recommends that a tubing set no more than 60cm (2ft.) in length with
no additional fittings be used for this purpose.
Calibration Procedure
1.
From the Main Menu Screen, click
to access the Setup-Calibration Screens.
Summary of Contents for Integritest Exacta
Page 2: ...Topics in this Manual...
Page 7: ......
Page 8: ...Using Help for the Integritest Exacta Instrument...
Page 11: ......
Page 12: ...Getting Started...
Page 20: ...Hardware Connections...
Page 32: ...First Log On...
Page 48: ...Starting Stopping the Instrument...
Page 58: ...Security Setup...
Page 69: ......
Page 70: ...Managing User Access...
Page 78: ...Instrument Setup...
Page 90: ...Filter Editor...
Page 101: ......
Page 102: ...Integritest Exacta Instrument Tests...
Page 109: ......
Page 110: ...Test Editor...
Page 122: ...Running a Test...
Page 131: ......
Page 132: ...Printing...
Page 139: ......
Page 140: ...Printer Maintenance and Troubleshooting...
Page 147: ......
Page 148: ...Data Management...
Page 171: ......
Page 172: ...Networking...
Page 175: ......
Page 176: ...Maintenance and Calibration...
Page 199: ......
Page 200: ...Troubleshooting...
Page 214: ...Error Messages...
Page 225: ......
Page 226: ...Instrument Repair...
Page 232: ...Warranty...
Page 235: ......
Page 236: ...Integritest Exacta Specifications...
Page 246: ...Interface Conventions...
Page 253: ......
Page 254: ...Contacting Millipore...
Page 262: ...References...
Page 263: ......
Page 265: ......
Page 266: ...Glossary...
Page 271: ......
Page 272: ...Index...