Integritest
Exacta Filter Integrity Test Instrument
Operating and Maintenance Manual
Hardware Connections
31
C.
Liquid
Trap
F.
Gas Inlet
*
optional item
Figure 6: Inlet and Outlet Tubing Connections
To Disconnect the Tubing
•
Push the collar of the fitting towards the instrument (see Figure 5).
•
Gently pull the tubing away from the instrument. (see Figure 5).
WARNING: Proper Installation is required
When installing the Integritest Exacta instrument, you must attach the outlet tubing assembly
and instrument valve to the top of the filter housing. The filter housing must be perpendicular
to the floor (Figure 7). For example, when using a T-line housing, you cannot attach the tubing
assembly to the housing inlet, which is located below the filter.
If the tubing assembly and instrument valve are not installed at the top of the housing, during
integrity testing residual fluid in the housing could be drawn back into the instrument when the
system is exhausted, and may cause damage to the instrument.
An improper installation of the outlet tubing assembly and instrument valve
to the filter housing will void your warranty.
Incorrect installation includes installing
the tubing assembly and the instrument valve below the filter or onto a filter housing that is not
upright.
CORRECT
INCORRECT
INCORRECT
Figure 7: Correct Filter Housing Assembly
Warning: Make sure all connections are tight before continuing.
Cutting the Tubing
1.
Cut the tubing squarely (See Figure 8).
Summary of Contents for Integritest Exacta
Page 2: ...Topics in this Manual...
Page 7: ......
Page 8: ...Using Help for the Integritest Exacta Instrument...
Page 11: ......
Page 12: ...Getting Started...
Page 20: ...Hardware Connections...
Page 32: ...First Log On...
Page 48: ...Starting Stopping the Instrument...
Page 58: ...Security Setup...
Page 69: ......
Page 70: ...Managing User Access...
Page 78: ...Instrument Setup...
Page 90: ...Filter Editor...
Page 101: ......
Page 102: ...Integritest Exacta Instrument Tests...
Page 109: ......
Page 110: ...Test Editor...
Page 122: ...Running a Test...
Page 131: ......
Page 132: ...Printing...
Page 139: ......
Page 140: ...Printer Maintenance and Troubleshooting...
Page 147: ......
Page 148: ...Data Management...
Page 171: ......
Page 172: ...Networking...
Page 175: ......
Page 176: ...Maintenance and Calibration...
Page 199: ......
Page 200: ...Troubleshooting...
Page 214: ...Error Messages...
Page 225: ......
Page 226: ...Instrument Repair...
Page 232: ...Warranty...
Page 235: ......
Page 236: ...Integritest Exacta Specifications...
Page 246: ...Interface Conventions...
Page 253: ......
Page 254: ...Contacting Millipore...
Page 262: ...References...
Page 263: ......
Page 265: ......
Page 266: ...Glossary...
Page 271: ......
Page 272: ...Index...