Millipore System Documentation Set
220
Operating and Maintenance Manual
Publication XITXSP121 P Rev.A
Table 27: Status Error Messages
Status
Error
No.
Status Window
Message
Short Description Detailed Description
1 MATH
ERROR
General
Failure
General error in operation
Used to convey errors occurring during operation
that include, improper set up parameters,
computational errors, etc.
2 INPUT
ERROR
Failed
Precondition
Check
Error that occurs at the initiation of a test. The
input parameters and the test & math dll‘s are
checked for validity prior to initiating the test.
This error condition will arise if a condition fails.
3 OUTPUT
ERROR
Failed
Postcondition
Check
Error that occurs at the completion of a test. The
output termination parameters and are checked
for validity at the completion of the test. This
error condition will arise if a condition fails.
4 INPUT
TOO
SMALL
Failed Param
Low Range
Check
Failed minimum boundary value for the input
value listed.
5 INPUT
TOO
HIGH
Failed Param
High Range
Check
Failed maximum boundary value for the input
value listed.
6 NETWORK
ERROR
RPC Failure
In network communications the RPC (Remote
Protocol Control) layer is the set of functions that
allow an IT Exacta hardware control software to
communicate with a remote User Interface (either
in an IT Exacta or on another computer). This
error message is generated when there is a
failure in this communications link and the two
are not connected.
7 NETWORK
UNAVAILABLE
RPC Service
Not Available
The Remote Communications Control package is
being used by another computer and unavailable
at this time.
8 GENERAL
ERROR
INSUFFI-
CIENT GAS
PRESSURE
Test Execution
General
Failure
General fault condition associated with an error
in the test protocol. Generally caused by incorrect
starting values being transmitted to the hardware
for either test operation or system configuration.
See the operations log for details.
Insufficient inlet gas pressure to run a test.
Summary of Contents for Integritest Exacta
Page 2: ...Topics in this Manual...
Page 7: ......
Page 8: ...Using Help for the Integritest Exacta Instrument...
Page 11: ......
Page 12: ...Getting Started...
Page 20: ...Hardware Connections...
Page 32: ...First Log On...
Page 48: ...Starting Stopping the Instrument...
Page 58: ...Security Setup...
Page 69: ......
Page 70: ...Managing User Access...
Page 78: ...Instrument Setup...
Page 90: ...Filter Editor...
Page 101: ......
Page 102: ...Integritest Exacta Instrument Tests...
Page 109: ......
Page 110: ...Test Editor...
Page 122: ...Running a Test...
Page 131: ......
Page 132: ...Printing...
Page 139: ......
Page 140: ...Printer Maintenance and Troubleshooting...
Page 147: ......
Page 148: ...Data Management...
Page 171: ......
Page 172: ...Networking...
Page 175: ......
Page 176: ...Maintenance and Calibration...
Page 199: ......
Page 200: ...Troubleshooting...
Page 214: ...Error Messages...
Page 225: ......
Page 226: ...Instrument Repair...
Page 232: ...Warranty...
Page 235: ......
Page 236: ...Integritest Exacta Specifications...
Page 246: ...Interface Conventions...
Page 253: ......
Page 254: ...Contacting Millipore...
Page 262: ...References...
Page 263: ......
Page 265: ......
Page 266: ...Glossary...
Page 271: ......
Page 272: ...Index...