Calibration Setup
Connect the Integritest Exacta inlet tubing to the instrument and to a regulated compressed
gas (nitrogen or air) source.
Set the inlet gas pressure to 100 psig (689kPa).
Note:
For an internal calibration, external tubing is not required and valves need not be
closed.
Calibration Procedure
1.
From the Main Menu Screen, click
to access the Setup-Calibration Screens.
2.
Click Tab F3 to open the Pressure Calibration Screen.
3.
Click
to start an internal calibration.
4.
The instrument will automatically begin to collect data for each calibration pressure.
5.
As the internal calibration terminates, pressure will be released.
6.
The calibration coefficients 1, 2, 3 and the regression coefficient for the secondary
transducer and pressure regulator will be updated automatically so they are matched
to the Primary transducer and appear on the screen. The information for the Primary
transducer is not adjusted.
7.
When the test is complete, print the report by clicking on the Print Icon
.
Summary of Contents for Integritest Exacta
Page 2: ...Topics in this Manual...
Page 7: ......
Page 8: ...Using Help for the Integritest Exacta Instrument...
Page 11: ......
Page 12: ...Getting Started...
Page 20: ...Hardware Connections...
Page 32: ...First Log On...
Page 48: ...Starting Stopping the Instrument...
Page 58: ...Security Setup...
Page 69: ......
Page 70: ...Managing User Access...
Page 78: ...Instrument Setup...
Page 90: ...Filter Editor...
Page 101: ......
Page 102: ...Integritest Exacta Instrument Tests...
Page 109: ......
Page 110: ...Test Editor...
Page 122: ...Running a Test...
Page 131: ......
Page 132: ...Printing...
Page 139: ......
Page 140: ...Printer Maintenance and Troubleshooting...
Page 147: ......
Page 148: ...Data Management...
Page 171: ......
Page 172: ...Networking...
Page 175: ......
Page 176: ...Maintenance and Calibration...
Page 199: ......
Page 200: ...Troubleshooting...
Page 214: ...Error Messages...
Page 225: ......
Page 226: ...Instrument Repair...
Page 232: ...Warranty...
Page 235: ......
Page 236: ...Integritest Exacta Specifications...
Page 246: ...Interface Conventions...
Page 253: ......
Page 254: ...Contacting Millipore...
Page 262: ...References...
Page 263: ......
Page 265: ......
Page 266: ...Glossary...
Page 271: ......
Page 272: ...Index...