![Millipore Integritest Exacta Operating And Maintenance Manual Download Page 115](http://html1.mh-extra.com/html/millipore/integritest-exacta/integritest-exacta_operating-and-maintenance-manual_1796911115.webp)
Millipore System Documentation Set
116
Operating and Maintenance Manual
Publication XITXSP121 P Rev.A
Entry
Value
required in
field?
Yes/No
Mandatory or
optional to modify
field when
programming a new
test?
Instructions
Notes
Filter Catalogue
Number
Yes Mandatory Location:
Tab F2
Select a filter
from the list.
Locate filters by either
scrolling through the list
using the arrow bars or
by using the “Quick
Search” feature. The
Quick Search feature
allows the user to type
in the first few
characters(up to four) of
the filter catalogue
number and will go
directly to the filter(s)
that correspond to
those characters.
Test Type
Yes Mandatory Location:
Tab F2
Select a test
from the list.
When a filter is chosen
the allowable tests for
that filter appear in the
Test Type dialog box.
Number of
Cartridges
Yes Optional
Default value is
commonly used.
Location:
Tab F2
Type in the
number of filters
that will be
tested with this
test program.
A default value of 1 will
appear. This only needs
to be modified if more
filters are being tested.
Housing Type
No Optional Location:
Tab F3
Type in any text
description that
is desired.
This is a text field that is
for descriptive
purposes.
Preset Upstream
Gas Volume
No Optional Location:
Tab F3
Click on the
check box and
then enter the
upstream
volume in the
dialog box.
This is used to bypass
the upstream gas
volume sizing, which
will not be performed if
this selection is chosen.
Instead the program will
use the Upstream Gas
Volume that is entered.
Summary of Contents for Integritest Exacta
Page 2: ...Topics in this Manual...
Page 7: ......
Page 8: ...Using Help for the Integritest Exacta Instrument...
Page 11: ......
Page 12: ...Getting Started...
Page 20: ...Hardware Connections...
Page 32: ...First Log On...
Page 48: ...Starting Stopping the Instrument...
Page 58: ...Security Setup...
Page 69: ......
Page 70: ...Managing User Access...
Page 78: ...Instrument Setup...
Page 90: ...Filter Editor...
Page 101: ......
Page 102: ...Integritest Exacta Instrument Tests...
Page 109: ......
Page 110: ...Test Editor...
Page 122: ...Running a Test...
Page 131: ......
Page 132: ...Printing...
Page 139: ......
Page 140: ...Printer Maintenance and Troubleshooting...
Page 147: ......
Page 148: ...Data Management...
Page 171: ......
Page 172: ...Networking...
Page 175: ......
Page 176: ...Maintenance and Calibration...
Page 199: ......
Page 200: ...Troubleshooting...
Page 214: ...Error Messages...
Page 225: ......
Page 226: ...Instrument Repair...
Page 232: ...Warranty...
Page 235: ......
Page 236: ...Integritest Exacta Specifications...
Page 246: ...Interface Conventions...
Page 253: ......
Page 254: ...Contacting Millipore...
Page 262: ...References...
Page 263: ......
Page 265: ......
Page 266: ...Glossary...
Page 271: ......
Page 272: ...Index...