Integritest
Exacta Filter Integrity Test Instrument
Operating and Maintenance Manual
Error Messages
221
Status
Error
No.
Status Window
Message
Short Description Detailed Description
9 UNDEFINED
TEST
Test Procedure
Not Found
At the time of execution, the test being asked for
is checked against the tests stored within the
internal database to ensure that the test is
properly defined and available for use. This error
indicates that the requested test procedure is not
found in the database test catalogue.
10 UNDEFINED
TEST
REVISION
Test Version
Not Found
Same as above, only the test version number is
the entity being checked.
11
TEST IN USE
Test Already
Running
Error condition flagged when the requested test is
already running on the IT Exacta hardware.
12
NO TEST
No Test
Available
Error condition that occurs at the initiation of a
test. If the test that is requested cannot be found
this error is generated. This condition occurs if an
archived test is used or if a stored test requests a
revision number that is different from the test
being requested.
13
NO TEST
Requested Test
Unavailable
Error condition that occurs at the initiation of a
test. If the test that is requested cannot be found
this error is generated. This condition occurs if an
archived test is used or if a stored test requests a
revision number that is different from the test
being requested.
14
NO DATA
Requested Test
Data
Unavailable
Error condition that occurs from the data
management function. The data being requested
cannot be found in the database.
15
GROSS LEAK Leak Detected
During the test, several checks are made (at every
data point) to determine the abrupt start of a
gross leak. The system is aborted if a gross leak
is found. A Gross leak can occur if tubing is
ruptured etc. or also in a bubble point or
enhanced bubble point test if the pressure is
increased too high and the filter is completely
blown dry.
16 TEST
TIME
EXCEEDED
Max Time
Exceeded
The maximum test time has been exceeded.
17
GROSS LEAK Max Loss
Exceeded
During each pressure decay the total pressure
drop is monitored, and if it exceeds half the test
pressure this error is shown.
Summary of Contents for Integritest Exacta
Page 2: ...Topics in this Manual...
Page 7: ......
Page 8: ...Using Help for the Integritest Exacta Instrument...
Page 11: ......
Page 12: ...Getting Started...
Page 20: ...Hardware Connections...
Page 32: ...First Log On...
Page 48: ...Starting Stopping the Instrument...
Page 58: ...Security Setup...
Page 69: ......
Page 70: ...Managing User Access...
Page 78: ...Instrument Setup...
Page 90: ...Filter Editor...
Page 101: ......
Page 102: ...Integritest Exacta Instrument Tests...
Page 109: ......
Page 110: ...Test Editor...
Page 122: ...Running a Test...
Page 131: ......
Page 132: ...Printing...
Page 139: ......
Page 140: ...Printer Maintenance and Troubleshooting...
Page 147: ......
Page 148: ...Data Management...
Page 171: ......
Page 172: ...Networking...
Page 175: ......
Page 176: ...Maintenance and Calibration...
Page 199: ......
Page 200: ...Troubleshooting...
Page 214: ...Error Messages...
Page 225: ......
Page 226: ...Instrument Repair...
Page 232: ...Warranty...
Page 235: ......
Page 236: ...Integritest Exacta Specifications...
Page 246: ...Interface Conventions...
Page 253: ......
Page 254: ...Contacting Millipore...
Page 262: ...References...
Page 263: ......
Page 265: ......
Page 266: ...Glossary...
Page 271: ......
Page 272: ...Index...