Millipore System Documentation Set
28
Operating and Maintenance Manual
Publication XITXSP121 P Rev.A
Pneumatic Connections
Diagram Key:
A.
Gas Inlet (see Gas Supply Requirements below)
B.
Exhaust
C.
From Housing
D.
To Housing
Figure 4: Side Panel
•
Gas Inlet (quick disconnect labeled “Air Inlet”)
•
Exhaust (tubing connector)
•
Outlet (dual quick disconnects labeled “From Housing” and “To Housing”)
Gas Supply Requirements
The instrument must be connected to a regulated supply of clean dry air (
>
-20 C dew point) or
nitrogen at 35–120 psig (2.4–8.5 bar).
Note:
To minimize the likelihood of water contamination, it is recommended that the inlet
gas be filtered with an external coalescing filter.
Gas Inlet Connection
To Gas Supply
•
Do not remove the gas inlet filter or damage to the instrument may result.
•
Connect a regulated supply of clean dry air or nitrogen, 35–120 psig (2.4–8.5
bar), to the inlet tubing assembly of the instrument using the ¼” NPTF side of the
gas filter.
•
Supply at least 15 psi (1.02 bar) more than the maximum pressure required to
perform the filter test.
To Instrument
•
Slide quick disconnect fitting of Integritest Exacta inlet tubing assembly onto the
connector stem labeled Air Inlet.
•
Grasp the rear of the connector and push it firmly against the connector stem. Do
not grasp the flared section when making a connection. This section must be able
to slide freely into place
•
Release it when you hear it click into place.
Summary of Contents for Integritest Exacta
Page 2: ...Topics in this Manual...
Page 7: ......
Page 8: ...Using Help for the Integritest Exacta Instrument...
Page 11: ......
Page 12: ...Getting Started...
Page 20: ...Hardware Connections...
Page 32: ...First Log On...
Page 48: ...Starting Stopping the Instrument...
Page 58: ...Security Setup...
Page 69: ......
Page 70: ...Managing User Access...
Page 78: ...Instrument Setup...
Page 90: ...Filter Editor...
Page 101: ......
Page 102: ...Integritest Exacta Instrument Tests...
Page 109: ......
Page 110: ...Test Editor...
Page 122: ...Running a Test...
Page 131: ......
Page 132: ...Printing...
Page 139: ......
Page 140: ...Printer Maintenance and Troubleshooting...
Page 147: ......
Page 148: ...Data Management...
Page 171: ......
Page 172: ...Networking...
Page 175: ......
Page 176: ...Maintenance and Calibration...
Page 199: ......
Page 200: ...Troubleshooting...
Page 214: ...Error Messages...
Page 225: ......
Page 226: ...Instrument Repair...
Page 232: ...Warranty...
Page 235: ......
Page 236: ...Integritest Exacta Specifications...
Page 246: ...Interface Conventions...
Page 253: ......
Page 254: ...Contacting Millipore...
Page 262: ...References...
Page 263: ......
Page 265: ......
Page 266: ...Glossary...
Page 271: ......
Page 272: ...Index...