Integritest
Exacta Filter Integrity Test Instrument
Operating and Maintenance Manual
Glossary
269
GLOSSARY
Bubble Point Pressure
Bubble point is based on the fact that liquid is held in the pores of a filter by surface tension and
capillary forces. The Bubble Point Pressure is the minimum pressure required to overcome these forces
and force liquid out of the pores of a membrane filter.
Bubble Point Test
A non-destructive filter integrity test that is based on the fact that liquid is held in the pores of a filter by
surface tension and capillary forces. The Bubble Point test detects the minimum pressure required to
overcome these forces and force liquid out of the pores of a membrane filter. The Bubble Point is most
critical for determining that a membrane is integral and meets specifications. This test procedure is also
useful for determining whether a correct cartridge has been installed in a filter housing.
Click
To press and release a mouse button quickly.
Default Password
Administrator
Default User Name
Administrator
Double-click
To rapidly press and release a mouse button twice without moving the mouse. Double clicking carries
out an action, such as starting an application.
Drag
To move an item on the screen by selecting the item and then pressing and holding down the mouse
button while moving the cursor. For example, you can move a window to another location on the
screen by dragging its title bar.
Diffusion
Diffusion is a migration of a solute from a region of high concentration to a region of lower
concentration. In most integrity tests the solute is air or nitrogen, which diffuses through a wetted
membrane.
Diffusion Test
A diffusion test is based on the fact that under pressure, compressed air or nitrogen flows through the
liquid that is held in the filter’s pores at differential pressures below the bubble point. In small area
filters, this airflow is very slow and often not measurable. But in large area filters, it is significant enough
to measure and provide a sensitive filter integrity test.
Downstream Side
The downstream side is the outlet, or sterile side of the filter through which fluid has passed.
Summary of Contents for Integritest Exacta
Page 2: ...Topics in this Manual...
Page 7: ......
Page 8: ...Using Help for the Integritest Exacta Instrument...
Page 11: ......
Page 12: ...Getting Started...
Page 20: ...Hardware Connections...
Page 32: ...First Log On...
Page 48: ...Starting Stopping the Instrument...
Page 58: ...Security Setup...
Page 69: ......
Page 70: ...Managing User Access...
Page 78: ...Instrument Setup...
Page 90: ...Filter Editor...
Page 101: ......
Page 102: ...Integritest Exacta Instrument Tests...
Page 109: ......
Page 110: ...Test Editor...
Page 122: ...Running a Test...
Page 131: ......
Page 132: ...Printing...
Page 139: ......
Page 140: ...Printer Maintenance and Troubleshooting...
Page 147: ......
Page 148: ...Data Management...
Page 171: ......
Page 172: ...Networking...
Page 175: ......
Page 176: ...Maintenance and Calibration...
Page 199: ......
Page 200: ...Troubleshooting...
Page 214: ...Error Messages...
Page 225: ......
Page 226: ...Instrument Repair...
Page 232: ...Warranty...
Page 235: ......
Page 236: ...Integritest Exacta Specifications...
Page 246: ...Interface Conventions...
Page 253: ......
Page 254: ...Contacting Millipore...
Page 262: ...References...
Page 263: ......
Page 265: ......
Page 266: ...Glossary...
Page 271: ......
Page 272: ...Index...