Millipore System Documentation Set
108
Operating and Maintenance Manual
Publication XITXSP121 P Rev.A
TEST PROCEDURES OF THE INTEGRITEST EXACTA INSTRUMENT
Diffusion Test Steps:
1.
System check
2.
Clear housing of excess fluid
•
Apply ½ of the test pressure to the housing.
3.
Determine upstream gas volume
4.
Pressurize the system to the test pressure
•
Allow the pressure to stabilize
5.
Run Pressure Decay Test
•
Measure pressure loss due to gas flow through the filter
6.
Determine diffusion flow
•
Calculate normalized diffusion constant (gas flow as a function of pressure).
•
When the diffusion constant (and therefore the diffusion flow rate) stabilizes, end
diffusion test.
Note:
The diffusion flow rate determination is stabilization driven, not event (i.e., Time)
driven.
Bubble Point Test Steps
1.
System check
2.
Clear housing of excess fluid
3.
Determine upstream volume
4.
Pressurize the system to the diffusion test pressure (or 80% of the bubble point
specification if the diffusion test pressure was not entered into the filter editor)
•
Allow the pressure to stabilize.
•
Increment pressure until bubble point is determined.
-
Increase pressure to 1 psi below the bubble point specification.
-
Increase pressure in 1 psi increments until 1 psi above the bubble point specification.
-
Increase pressure in increments proportional to 2.5 psi (1 psi for IPA tests) until bubble
point is reached. This proportionally reduces the step size as pressure gets further away
from the minimum bubble point specification.
-
Increase pressure in various increments above the bubble point until the test is complete.
5.
Calculate decay
•
Calculate pressure loss after each pressure-decay period.
6.
Continue testing
•
The instrument determines changes in flow by an Intersection of Tangents method.
The intersection between the diffusion curve and a linear fit of the flowrates above
the bubble point is defined as the bubble point.
Enhanced Bubble Point Test Steps
The
Enhanced Bubble Point
test is a combination of the diffusion and bubble point tests.
1.
System check
2.
Clear housing of excess fluid
Summary of Contents for Integritest Exacta
Page 2: ...Topics in this Manual...
Page 7: ......
Page 8: ...Using Help for the Integritest Exacta Instrument...
Page 11: ......
Page 12: ...Getting Started...
Page 20: ...Hardware Connections...
Page 32: ...First Log On...
Page 48: ...Starting Stopping the Instrument...
Page 58: ...Security Setup...
Page 69: ......
Page 70: ...Managing User Access...
Page 78: ...Instrument Setup...
Page 90: ...Filter Editor...
Page 101: ......
Page 102: ...Integritest Exacta Instrument Tests...
Page 109: ......
Page 110: ...Test Editor...
Page 122: ...Running a Test...
Page 131: ......
Page 132: ...Printing...
Page 139: ......
Page 140: ...Printer Maintenance and Troubleshooting...
Page 147: ......
Page 148: ...Data Management...
Page 171: ......
Page 172: ...Networking...
Page 175: ......
Page 176: ...Maintenance and Calibration...
Page 199: ......
Page 200: ...Troubleshooting...
Page 214: ...Error Messages...
Page 225: ......
Page 226: ...Instrument Repair...
Page 232: ...Warranty...
Page 235: ......
Page 236: ...Integritest Exacta Specifications...
Page 246: ...Interface Conventions...
Page 253: ......
Page 254: ...Contacting Millipore...
Page 262: ...References...
Page 263: ......
Page 265: ......
Page 266: ...Glossary...
Page 271: ......
Page 272: ...Index...