Enhanced Bubble Point
The Enhanced Bubble Point test is a combination of a diffusion test followed by a bubble point test to
determine if a filter is integral. The Enhanced Bubble Point test is the test recommended by Millipore for
most applications. Exceptions to this are;
•
low area devices which do not have diffusion specifications
•
large multi-cartridge systems where it is difficult to determine the bubble point.
HydroCorr Test
The HydroCorr test is a water-based integrity test for hydrophobic filters.
HydroCorr Test Initial Gas Volume
The volume of gas located on the upstream side of a HydroCorr test filter before test pressure is
applied. This volume must be maintained within the range defined by the Minimum and Maximum Gas
Volumes. If the actual initial gas volume is outside of this range, the accuracy of the test could be
compromised. An error message will result and the test will stop.
HydroCorr Test Time
The HydroCorr test time defines the duration of the test. Unlike a Diffusion test, the HydroCorr test will
run for the entire time programmed here.
Hydrophilic
A term used to describe molecules or molecular groups that are attracted to water and other polar
solvents.
Hydrophobic
A term used to describe molecules or molecular groups that mix poorly with water and other polar
solvents.
Integrity Test
Non-destructive, non-contaminating test used to determine whether or not a filter has a defect capable
of passing particles that should be retained.
Legend
Legends may include any form of test-specific data, which a user might wish to associate with the testing
and reporting process:
•
Lot Numbers
•
Serial Numbers
Mandatory
If data is defined as mandatory, a test cannot be run without the designated information.
Membrane Filter
A matrix that removes particles larger than the specified pore size.
Summary of Contents for Integritest Exacta
Page 2: ...Topics in this Manual...
Page 7: ......
Page 8: ...Using Help for the Integritest Exacta Instrument...
Page 11: ......
Page 12: ...Getting Started...
Page 20: ...Hardware Connections...
Page 32: ...First Log On...
Page 48: ...Starting Stopping the Instrument...
Page 58: ...Security Setup...
Page 69: ......
Page 70: ...Managing User Access...
Page 78: ...Instrument Setup...
Page 90: ...Filter Editor...
Page 101: ......
Page 102: ...Integritest Exacta Instrument Tests...
Page 109: ......
Page 110: ...Test Editor...
Page 122: ...Running a Test...
Page 131: ......
Page 132: ...Printing...
Page 139: ......
Page 140: ...Printer Maintenance and Troubleshooting...
Page 147: ......
Page 148: ...Data Management...
Page 171: ......
Page 172: ...Networking...
Page 175: ......
Page 176: ...Maintenance and Calibration...
Page 199: ......
Page 200: ...Troubleshooting...
Page 214: ...Error Messages...
Page 225: ......
Page 226: ...Instrument Repair...
Page 232: ...Warranty...
Page 235: ......
Page 236: ...Integritest Exacta Specifications...
Page 246: ...Interface Conventions...
Page 253: ......
Page 254: ...Contacting Millipore...
Page 262: ...References...
Page 263: ......
Page 265: ......
Page 266: ...Glossary...
Page 271: ......
Page 272: ...Index...