Millipore System Documentation Set
208
Operating and Maintenance Manual
Publication XITXSP121 P Rev.A
Gross Leak detection prior to sizing
Prior to sizing, the system detects gross leaks by comparing the measured pressure drop
against the pressure drop that would be expected if a leak flowrate is set at twice the filter
diffusion specification. The pressure drop at this flowrate can be computed using the standard
diffusion equation:
dP/dt
limit
= 2* Q
spec
*14.7 / V
However, since the actual volume is unknown, the table below lists the assumed volumes.
Table 25: Assumed Volume Used for Leak Detection
Filter Classification
Q spec
(mL/min)
Assumed volume
(mls)
Hydrophilic
> 500
3600
3 < Q <
500
1800
<
3
425
Hydrophobic
> 500
NA
3 < Q <
500
1800
<
3
152
HydroCorr >0.4
150
<=
0.4
25
<0.1
25
Listed below is an example of how this criterion would be applied.
A Millipak would have a default Q
spec
of 3 mL/min. and an assumed volume of 425 mL. The
maximum change in pressure over time would be determined as follows:
dP/dt
limit
= [2 * (3 mL/min) * (14.7 psi)]/[425 mL]
Based on this example, any pressure drop exceeding 0.21 psi/min would prompt this
message.
Summary of Contents for Integritest Exacta
Page 2: ...Topics in this Manual...
Page 7: ......
Page 8: ...Using Help for the Integritest Exacta Instrument...
Page 11: ......
Page 12: ...Getting Started...
Page 20: ...Hardware Connections...
Page 32: ...First Log On...
Page 48: ...Starting Stopping the Instrument...
Page 58: ...Security Setup...
Page 69: ......
Page 70: ...Managing User Access...
Page 78: ...Instrument Setup...
Page 90: ...Filter Editor...
Page 101: ......
Page 102: ...Integritest Exacta Instrument Tests...
Page 109: ......
Page 110: ...Test Editor...
Page 122: ...Running a Test...
Page 131: ......
Page 132: ...Printing...
Page 139: ......
Page 140: ...Printer Maintenance and Troubleshooting...
Page 147: ......
Page 148: ...Data Management...
Page 171: ......
Page 172: ...Networking...
Page 175: ......
Page 176: ...Maintenance and Calibration...
Page 199: ......
Page 200: ...Troubleshooting...
Page 214: ...Error Messages...
Page 225: ......
Page 226: ...Instrument Repair...
Page 232: ...Warranty...
Page 235: ......
Page 236: ...Integritest Exacta Specifications...
Page 246: ...Interface Conventions...
Page 253: ......
Page 254: ...Contacting Millipore...
Page 262: ...References...
Page 263: ......
Page 265: ......
Page 266: ...Glossary...
Page 271: ......
Page 272: ...Index...