Chapter 6 Trigger System
101
Trigger Option
Setting
Description
CH3
Set CH3 as trigger signal source
CH4
Set CH4 as trigger signal source
Polarity
Positive
Trigger on setting the positive pulse width of signals
Negative
Trigger on setting the negative pulse width of signals
Trigger Condition
<
T
Trigger when the signal pulse width is smaller than pulse width T
>
T
Trigger when the signal pulse width is greater than pulse width T
=
T
Trigger when the signal pulse width is equal to pulse width T
≠T
Trigger when the signal pulse width is not equal to pulse width T
Trigger Pulse
Width
8ns~10s
Set the trigger pulse width
Notes:
Conditions of greater than, smaller than, equal to or not equal to indicating that the error is 6%.
Trigger steps of positive polarity pulse width: (taking CH1 as an example)
Summary of Contents for Smart Series
Page 1: ......
Page 2: ...Version Info Version Date Remarks V2 0 2019 10 ...
Page 25: ...Chapter 2 Quick Start Guide of Oscilloscope 11 2 3 Side Panel Figure 2 2 Side Panel ...
Page 30: ...16 Figure 2 5 Oscilloscope Interface Display ...
Page 51: ...Chapter 3 Operate the Oscilloscope with Buttons 37 Figure 3 1 Button Areas ...
Page 61: ...Chapter 4 Horizontal System 47 Figure 4 1 Horizontal system ...
Page 82: ...68 Figure 5 4 Channel Open Close and Switching ...
Page 88: ...74 Figure 5 7 DC Coupling Figure 5 8 AC Coupling ...
Page 126: ...112 Figure 6 21 Trigger Level Adjustment Trigger pulse width time can be set as 8ns 10s ...
Page 150: ...136 Figure 7 2 Pull up Menu Figure 7 3 All Measurements ...
Page 161: ...Chapter 7 Analysis System 147 Figure 7 10 Cursor Measurement Description ...
Page 176: ...162 Figure 8 5 Delete Reference Files CSV files CSV file structure ...
Page 190: ...176 Figure 9 4 Spectrum Amplitude as V Hz Adjust FFT waveforms Waveform position ...
Page 324: ...310 Figure 14 6 Android APP Figure 14 7 Successful Connection of Android APP ...