Chapter 11 Sampling System
207
Sampling rate = record length/acquisition time
Generally, the oscilloscope acquisition time is exactly the display time on the current entire screen (current
horizontal time base×14).
For example, if the oscilloscope has the memory depth of 700K, the sampling rate of 1GSa/s, and the horizontal
time base of 50us/div, the acquisition time is 700us, which is 50us/div×14div.
However, when the fast time base (below 20 ns) or the record length is set to a fixed value, the oscilloscope
acquisition time does not necessarily represent the display time on the current entire screen.
For example, if the oscilloscope has the memory depth of 700K, the sampling rate of 1GSa/s, and the horizontal
time base of 20ns, the acquisition time is 700ns, which is 2.5 times of the current display time on the entire screen.
Or, if the memory depth is 140K (fixed value), the sampling rate is 1GSa/s, and the horizontal time base is 1us, the
acquisition time is 140us, which is 10 times of the current display time on the entire screen.
For a single channel in a channel pair, the maximum sampling rate of the Smart Series oscilloscope is 1GSa/s.
Summary of Contents for Smart Series
Page 1: ......
Page 2: ...Version Info Version Date Remarks V2 0 2019 10 ...
Page 25: ...Chapter 2 Quick Start Guide of Oscilloscope 11 2 3 Side Panel Figure 2 2 Side Panel ...
Page 30: ...16 Figure 2 5 Oscilloscope Interface Display ...
Page 51: ...Chapter 3 Operate the Oscilloscope with Buttons 37 Figure 3 1 Button Areas ...
Page 61: ...Chapter 4 Horizontal System 47 Figure 4 1 Horizontal system ...
Page 82: ...68 Figure 5 4 Channel Open Close and Switching ...
Page 88: ...74 Figure 5 7 DC Coupling Figure 5 8 AC Coupling ...
Page 126: ...112 Figure 6 21 Trigger Level Adjustment Trigger pulse width time can be set as 8ns 10s ...
Page 150: ...136 Figure 7 2 Pull up Menu Figure 7 3 All Measurements ...
Page 161: ...Chapter 7 Analysis System 147 Figure 7 10 Cursor Measurement Description ...
Page 176: ...162 Figure 8 5 Delete Reference Files CSV files CSV file structure ...
Page 190: ...176 Figure 9 4 Spectrum Amplitude as V Hz Adjust FFT waveforms Waveform position ...
Page 324: ...310 Figure 14 6 Android APP Figure 14 7 Successful Connection of Android APP ...