100
When there is low frequency noise in the waveform, stable edge triggering can be obtained using LF rejection
coupling.
5)
NoiseRej. (Noise Rejection Coupling) - Noise rejection can add extra hysteresis to the trigger circuit. By
increasing the trigger hysteresis band, the possibility of noise triggering can be reduced. But it also reduces the
trigger sensitivity, so triggering the oscilloscope requires a slightly larger signal.
Note:
Trigger coupling is independent of channel coupling.
6.3 Pulse Width Trigger
The trigger happens when the trigger signal pulse width (8ns~10s, the trigger type indication icon at the top of the
screen is
) reaches the set condition and the signal voltage reaches the set trigger level. Pulse width trigger
menu is shown in the following table:
Trigger Option
Setting
Description
Trigger Source
CH1
Set CH1 as trigger signal source
CH2
Set CH2 as trigger signal source
Summary of Contents for Smart Series
Page 1: ......
Page 2: ...Version Info Version Date Remarks V2 0 2019 10 ...
Page 25: ...Chapter 2 Quick Start Guide of Oscilloscope 11 2 3 Side Panel Figure 2 2 Side Panel ...
Page 30: ...16 Figure 2 5 Oscilloscope Interface Display ...
Page 51: ...Chapter 3 Operate the Oscilloscope with Buttons 37 Figure 3 1 Button Areas ...
Page 61: ...Chapter 4 Horizontal System 47 Figure 4 1 Horizontal system ...
Page 82: ...68 Figure 5 4 Channel Open Close and Switching ...
Page 88: ...74 Figure 5 7 DC Coupling Figure 5 8 AC Coupling ...
Page 126: ...112 Figure 6 21 Trigger Level Adjustment Trigger pulse width time can be set as 8ns 10s ...
Page 150: ...136 Figure 7 2 Pull up Menu Figure 7 3 All Measurements ...
Page 161: ...Chapter 7 Analysis System 147 Figure 7 10 Cursor Measurement Description ...
Page 176: ...162 Figure 8 5 Delete Reference Files CSV files CSV file structure ...
Page 190: ...176 Figure 9 4 Spectrum Amplitude as V Hz Adjust FFT waveforms Waveform position ...
Page 324: ...310 Figure 14 6 Android APP Figure 14 7 Successful Connection of Android APP ...