5-18
MICRONIX USA, LLC
www.micronixusa.com
Rev: 3.02
Clear Errors
During Motion
Real-time
Program
Global
Set
Read
Set
Read
Set
Read
Set
Read
Command
Description:
This command is used to clear all error messages without reading
them.
Returns:
A read operation cannot be used with this command.
Syntax:
nCER
–
Standard syntax
0CER
–
All axes clear error messages
Parameter
Description:
n[int]
–
Axis number
Parameter
Range:
n
–
0 to 99
Related Commands: ERR
Example:
1CER
|
Axis 1, clear error messages
–
0CER
|
All axes, clear error messages
Summary of Contents for MMC-200 Series