Page 93
Data Processing
thermoIMAGER TIM Connect
Under
Extended settings
, a
Holdoff time
can be en-
tered. A new alarm condition is not taken into account during
this set time.
If the checkmark is enabled by
Allow event grabbing
(and alarm)
while playing
, the event grabber is consid-
ered when playing a Ravi.
Create snapshot on event
will create a snapshot after
every event which will appear in the snapshot history and in
the event grabber (frozen image) window.
The triggered snapshot for the event grabber can be found under menu
View
,
Windows
and
Event grab-
ber (frozen Image).
Measurement areas can be created and arranged in the live IR image as well as in the event grabber (frozen
image). If you want to add an measurement area in the event grabber (frozen image), this must be set under
the
Configuration
menu and
Measure areas
. Under
Measure area
, the check mark
Use in event
grabber
must be selected.
As a template for the event grabber, you can use the sample layout
Event grabber
.