Page 117
Line Scanner Mode
thermoIMAGER TIM Connect
If activated
Avoid re-triggering during scan
will not allow to run a new line scan cycle while a previ-
ous cycle trigger by recent signal is not finished.
Using Avoid temperature analyzing, detecting alarm states and PIF operation during scan, the three opera-
tions mentioned are performed only at the end of the scan. This is very helpful for performance reasons when
a very high sampling rate (e.g. 1000 Hz, TIM 1M) is set.
i
For external triggering it is necessary to do modifications of the standard configuration on the Process
Interface (PIF). You will find more details, see Chap.
4.4
.
Using the Self triggered option it is possible to use a triggering signal provided by software to control arrang-
ing of the scanned lines. Number of lines being reloaded after this signal is defined under Lines and History
[sec], see Chap.
9.3.1
.