Page 40
Software Configuration
thermoIMAGER TIM Connect
4.3.7
Changing the Temperature Range
Using the menu
Tools
,
Configuration
and
Device
the most suitable
Temperature range
for the pro-
cess can be set. The thermoIMAGER TIM features different temperature ranges depending on the device.
Fig. 38 View Temperature Range
Alternatively, the temperature range can also be changed via the status bar in the PIX Connect software win-
dow. To do this, make a right-click on the temperature range.
A useful function is the adjustable, extended temperature range under Extended. The temperature range
150 °C ... 900 °C can be extended from 20 °C. The accuracy between 20 °C ... 150 °C is no longer exact. This
function is used to display an IR image with or without temperatures between 20 °C ... 150 °C. This is espe-
cially useful if an IR image is to be displayed between the hot processes.
i
The extended temperature range is usable for all TIM 160 /2xx /4xx /640 cameras, calibrated from
March 2017 onwards and for all TIM 8 TIM 40 cameras.
With
High resolution temperatures (if available)
, the temperature display has two decimal
places instead of one decimal place. This function is only available for certain camera models.