MTV1000 Ventilator System
69
User’s Manual
7.2.2.
V-SIMV(Volume based Synchronized Intermittent Mandatory
Ventilation)
V-SIMV is a mode that controls ventilation based on volume according to spontaneous
ventilation of the patient. In other words, this mode combined control and pressure
support/ spontaneous function allows for preset mandatory breaths synchronized with
the patient’s breathing.
Spontaneous ventilation is done using configured time period or mechanical ventilation is
paralleled based on air volume if there is no spontaneous breathing.
The spontaneous/pressure supported breaths are defined by the setting for Pressure
Support.
WARNING
Mechanical ventilation is paralleled based on volume according to spontaneous ventilation
or in case of no spontaneous ventilation based on time period configured.
SIMV mode is a mode in which Mandatory Breath and Spontaneous Breath are mixed.
SIMV mode guarantees one mandatory breath per SIMV cycle. This mandatory breath is
either patient initiated mandatory (PIM or assisted mandatory) breath or ventilator
initiated mandatory (VIM or controlled mandatory) breath.
As in the figure below, each SIMV breath cycle is divided into two parts. One is
Mandatory Interval (Tm) and another is Spontaneous Interval (Ts). Once PIM breath is
delivered, mandatory interval ends and spontaneous interval (Ts) begins.
If PIM breath does not occur until mandatory interval ends, VIM breath is provided to the
patient as soon as mandatory interval ends and spontaneous interval begins.