MTV1000 Ventilator System
54
User’s Manual
6.11.
System menu
System menu is composed of “SETUP” menu that configures operating conditions of the
ventilator, “Date” menu that decides date and time, and “Calibration” menu that performs
various calibrations, and “Patient” menu that edit patient information.
6.11.1.
SETUP
Item
Range
Unit
STEP
Description
BWF
5~15
mL/kg
1
Configure tidal volume per Kg of the patient’s
body weight(B/W).
O
2
Cell
ON / OFF
-
-
Configure the use of O
2
cell, whether to enable
or disable the sensor.
ON : O
2
gas is used when O
2
cell is normal.
OFF : This is when there is no O
2
gas, lifespan of
O
2
cell has expired, or O
2
cell can not be used
for a different reason. There is no influence on
supply of O
2
gas.
Trend Init
ON / OFF
-
-
Initialize trend data.
BTPS*
OFF
-
-
BTPS is turned OFF.
Manual DRY
-
-
BTPS is turned on in the condition that
humidifier is not used.
Manual Humid
-
-
BTPS is turned on in the condition that
humidifier is used.
Sound Vol.
10 ~ 100
%
10
Decide alarm sound volume.
ALTITUDE
-100 ~ 5000
m
20
Configure altitude above sea level.
Air Temp
0 ~ 50
Celsius
1
Configure environmental temperature.
Proximal
ON / OFF
-
-
Configure the use of proximal pressure sensor.
OFF : Proximal Pressure Sensor Off
ON : Proximal Pressure Sensor On