MTV1000 Ventilator System
26
User’s Manual
4.2.
Internal Battery
Internal battery is used to supply power to the product when the patient is moving.
Battery is automatically charged when external commercial power is supplied. Once
commercial power supply is terminated or power supply is suspended accidentally, power
is automatically switched to internal battery. Remaining battery status is indicated on the
LCD.
WARNING
Battery must be charged before use. Replace the battery if it is considered inadequate.
If battery gets low, measurement information relating ventilation may result in errors. If
remaining battery is low, connect to commercial power.
Be careful about polarity when replacing battery. Connect (+) to red and (-) to black. If
polarity is mistaken, it may damage the product.
Do not short battery terminal. (Battery may explode and cause harm on human body)
Use rated voltage and current. (10.8V, 4400mA)
In case of abnormality in battery, contact [Customer Service team] to receive necessary
follow-up.
CAUTION
Fuse is internalized in the product.
Time Lag Type, 250V, 6.3A
Internal battery can be influenced by the number of discharges during use, degree of
discharge, surrounding temperature, and charging voltage. Battery can ordinarily be used
for two years.