MTV1000 Ventilator System
20
User’s Manual
INS Low(optional)
OFF, 0.0 ~ 14.9 %
RESP High(optional)
3 ~ 150 rpm
RESP Low(optional)
2 ~ 149 rpm
3.3.6.
External Specification
Main body size
25 x 15.6 x 37.1 (Width x Depth x Height) cm
Main body mass
8 Kg
Inspiratory limb connector
ISO 22mm conical male
Expiratory limb connector
ISO 22mm conical male
Oxygen supply inlet
DISS male / female
3.3.7.
Environmental Specification
Operating Temperature range
10 to 40
℃
(50 ~ 104
℉
)
Operating Atmosphere Pressure range
700 to 1060 mbar (10.2 ~ 15.4 psi)
Operating Altitude range
Up to 3000 m (9,842 ft)
Operating Humidity range
0 to 95 %
Storage Temperature range
-20 to 50
℃
Storage Atmosphere Pressure range
500 to 1060 mbar (7.3 ~ 15.4 psi)
Storage Altitude range
Up to 6560 m (20,000 ft)
Storage Humidity range
10 to 90 %
O
2
supply pressure range
2.56 ~ 7.3 Kgf/
㎠
or 2.4 ~ 6.9 bar
or 35 ~ 90 psig
International Standards
This product selected and acquired the following international standards.
EN60601-1:1990
Electric Safety
EN60601-1-1:2001
Electromechanical Safety
EN60601-1-2:2001
Electromagnetic Compatibility Requirement and tests
IEC 60601-2-12:2001
Particular requirements for the safety of lung ventilators for medical
use