MTV1000 Ventilator System
66
User’s Manual
7.2.
Ventilation Modes
7.2.1.
V-ACV(Volume Assist Control Ventilation)
Volume control mode is based on mechanical ventilation method and the ventilator
delivers configured tidal volume to the patient.
VACV mode is a mode that controls mechanical ventilation of the patient based on air
volume. It controls inspiratory and expiratory times by force
In Assist/Control mode, Assisted/Controlled Mandatory Breath is provided regardless of
period once the device detects intention of the patient to breathe. If there is no intention
to breathe, air is supplied based on Assisted/Controlled Mandatory Breath period.
Item
Range
Unit
Step
Description
B/W
10~150
kg
1
Configure patient’s body weight.
V
TIDAL
50 ~ 2500
mL
5
Air volume of the patient at a time
R
ATE
2 ~ 60
BPM
1
Configure number of inspirations per minute.
P
EEP
0 ~ 50
cmH
2
O
1
Configure PEEP (base pressure of patient).
O
2
21 ~ 100
%
1
Configure O
2
(oxygen) concentration of gas
supplied to the patient.
TI
0.2 ~ 9.9
sec
0.1
Configure inspiration time (patient).
E
n
SENS
10 ~ 80
%
5
Configure sensitivity with which each breath is
completed by the patient using % expiration.
TRIGGER
TYPE
Pressure/
Flow
-
-
Configure spontaneous ventilation detection
method as pressure or flow type.