MTV1000 Ventilator System
28
User’s Manual
4.2.2.
Internal Battery Replacement Method
No.
Figure
Description
1
Unscrew the six screws on both sides of the
battery cover.
2
After removing battery cover, separate battery
pack.
3
There are two batteries. Replace the wanted
battery and assemble in a reverse sequence.
WARNING
Place battery connector in the correct position when replacing battery. If the connector is
misplaced, it may damage the product.
CAUTION
All bolts are tightened clockwise and loosened counterclockwise.