MTV1000 Ventilator System
58
User’s Manual
6.12.
Alarm Setting
1/2 page shows alarms relating to ventilation. 2/2 page shows SpO
2
and EtCO
2
monitoring
alarms. Set appropriate alarm limit using touch and encoder.
Page
Item
Range
Unit
STEP
Description
1/2
VT
E
10 ~ 2500
,OFF
mL
10
Configure upper / lower limits for volume
per breath.
V
E MIN
0.0 ~ 50
LPM
0.1
Configure upper / lower limits for volume
per minute.
R
ATE
0 ~ 150
BPM
1
Configure upper / lower limits for number
of breaths per minute.
P
AW
0 ~ 120,
OFF
cmH
2
O 1
Configure upper / lower limits for
pressure.
O
2
0 ~ 100,
OFF
%
1
Configure upper / lower limits for oxygen
concentration.
A
IR LEAK
10 ~ 500,
OFF
mL
10
Configure difference between inspiratory
and expiratory volumes.