MTV1000 Ventilator System
77
User’s Manual
Pressure Regulation Operation at Compliance Change of Patient
Item
Range
Unit
Step
Description
B/W
10~150
kg
1
Configure pressure supplied during inspiration.
V
TIDAL
50 ~ 2500 mL
5
Air volume of the patient at a time
T
INSP
0.2 ~ 9.9
sec
0.1
Configure patient’s inspiration time.
Trise
Fast/Mediu
m/Slow
-
-
Configure arrival time to peak flow or peak
pressure.
R
ATE
2 ~ 60
BPM
1
Configure number of inspirations per minute.
P
EEP
0 ~ 50
cmH
2
O 1
Configure PEEP (base pressure of patient).
O
2
21 ~ 100
%
1
Configure O
2
(oxygen) concentration supplied to the
patient.
TRIGGER
Type
Pressure/
Flow
-
-
Configure
spontaneous
ventilation
detection
method as pressure or flow type.
F(P)
TRIG
OFF, 0.5 ~
20
cmH
2
O/
LPM
0.5
Configure sensitivity for detecting spontaneous
ventilation by the patient.
P LIMIT
5 ~ 80
cmH
2
O 1
Configure pressure limit supplied to the patient
during spontaneous ventilation.
MASK
ON/OFF
-
-
Configure mask on / off.
ACCEPT
-
-
-
Execute P-SIMV Mode.
CANCEL
-
-
-
Move back to the previous menu.