MTV1000 Ventilator System
19
User’s Manual
3.3.4.
Useful Function Description
Function
Description
Inspiration pause
Measures patient’s lung resistance, compliance, elasticity, time
constant
Manual Inspiration
Delivers one mandatory breath
Touch-Key lock
Disable touch-key function
Graph freeze
Freezes waveform & loop graphs
Alarm silence
Turns off alarm sound for 2 minutes
Alarm reset
Clear active alarms
Sound Volume
Controls alarm sound volume
3.3.5.
Alarm Specification
Item
Range
PRESSURE HIGH
OFF, 0 cmH
2
O ~ 120 cmH
2
O
PRESSURE LOW
OFF, 0 cmH
2
O ~ 110 cmH
2
O
VTIDAL High
OFF, 20 ml ~ 2500 ml
VTIDAL LOW
OFF, 10 ml ~ 2400 ml
O2 HIGH
OFF, 0% ~ 100%
O2 LOW
OFF, 0% ~ 80%
RATE HIGH
2 ~ 150 BPM
RATE LOW
0 ~ 148 BPM
VE MIN HIGH
0.5 ~ 50 LPM
VE MIN LOW
0.0 ~ 49.0 LPM
O2 Pressure Fail
-
AIR LEAK
10 ml ~ 500 ml, OFF
APNEA
2 ~ 60 sec
VENT INOP
-
O2 CELL ERROR
-
CIRCUIT OPEN
-
BATTERY MODE
-
LOW BATTERY
DC power loss
COM ERROR
-
SpO2 High(optional)
OFF, 52 ~ 100%
SpO2 Low(optional)
51 ~ 100%
PR High(optional)
26 ~ 250 bpm
PR Low(optional)
25 ~ 249 bpm
EtCO2 High(optional)
0.0 ~ 15.0 %
EtCO2 Low(optional)
OFF, O.O ~ 14.9%
INS High(optional)
0.0 ~ 15.0 %