MTV1000 Ventilator System
70
User’s Manual
In SIMV mode, SIMV Breath Cycle is determined by the number of breaths configured. If
Respiratory Rate is configured as 12 times, SIMV Breath Cycle is 5 seconds. Mandatory
Interval is configured as smaller value between 10 seconds and 0.6 x SIMV Breath Cycle.
Mandatory Interval may be shortened depending on PIM Breath.
Item
Range
Unit
Step Description
B/W
10~150
kg
1
Configure patient’s body weight.
V
TIDAL
50 ~ 2500
mL
5
Volume of the patient at a time
R
ATE
2 ~ 60
BPM
1
Configure number of inspirations per minute.
APNEA
2 ~ 60
sec
1
Configure time to convert to Apnea Mode if
patient shows no ventilation.
P
EEP
0 ~ 50
cmH
2
O 1
Configure PEEP (base pressure of patient).
O
2
21 ~ 100
%
1
Configure O
2
(oxygen) concentration of gas
supplied to the patient.
TI
0.2 ~ 9.9
sec
0.1
Configure inspiration time (patient).
Trise
Fast/Mediu
m/Low
-
-
Configure arrival time to peak flow or peak
pressure.
E
n
SENS
10 ~ 80
%
5
Configure sensitivity with which each breath is
completed by the patient using % expiration.
E
x
SENS
10 ~ 80
%
10
Configure spontaneous ventilation detection
method as pressure or flow type.
TRIGGER
TYPE
PRESSURE/
FLOW
-
-
Configure sensitivity for detecting spontaneous
ventilation by the patient. Choose between
pressure or flow type.
F(P)
TRIG
OFF, 0.5 ~
20
cmH
2
O/
LPM
0.5
Configure sensitivity to detect spontaneous
ventilation by the patient.
PS
0 ~ 60
cmH
2
O 1
Configure pressure supplied to the patient during
spontaneous ventilation.