MTV1000 Ventilator System
79
User’s Manual
Item
Range
Unit
Step
Description
B/W
10 ~150
kg
1
Configure patient’s body weight.
P
EEP
0 ~ 50
cmH
2
O
1
Configure PEEP (base pressure of patient).
O
2
21 ~ 100
%
1
Configure O
2
(oxygen) concentration supplied to the
patient.
EnSENS
10 ~ 80
%
5
Configure sensitivity with which each breath is
completed by the patient using % expiration.
E
x
SENS
10 ~ 80
%
5
Configure sensitivity with which each breath is
completed by the patient using % inspiration.
TRIGGER
TYPE
Pressure/
Flow
-
-
Configure spontaneous ventilation detection method
as pressure or flow type.
F(P)
TRIG
OFF, 0.5 ~
20
cmH
2
O/
LPM
0.5
Configure sensitivity for detecting spontaneous
ventilation by the patient.
PS
0 ~ 60
cmH
2
O
1
Configure pressure supplied to the patient during
spontaneous ventilation.
Trise
Fast/Mediu
m/Slow
-
-
Configure arrival time to peak flow or peak
pressure.
APNEA
2 ~ 60
sec
1
Configure time to convert to Apnea Mode if patient
shows no ventilation.
APNEA
BACKUP
TYPE
V-ACV/
P-ACV
-
-
Configure V-ACV or P-ACV when converting to
Apnea Mode.
APNEA
PINSP
5 ~ 80
cmH
2
O
1
Configure pressure supplied during inspiration.
APNEA
V
T
50 ~ 2500 mL
5
Volume of the patient at a time
MASK
ON/OFF
-
-
Configure mask on / off.
ACCEPT
-
-
-
Execute SPONT+ Mode.
CANCEL
-
-
-
Move back to the previous menu.