MTV1000 Ventilator System
49
User’s Manual
6.7.
Monitoring menu
“Monitoring” menu shows all measuring information comprehensively.
Table 1 shows details on ventilation. Table 2 shows details on SpO
2
and EtCO
2
.
Monitoring list according to ventilation mode is as below.
Item
Unit
Description
Paw
cmH
2
O
Airway pressure
P PEAK
cmH
2
O
Maximum inspiratory pressure
P MEAN
cmH
2
O
Mean airway pressure
P PAUSE
cmH
2
O
Pause inspiratory pressure
PEEP
cmH
2
O
Positive end expiratory pressure
Auto PEEP
cmH
2
O
Difference between measured PEEP and configured PEEP
VI TIDAL
mL
Inspiratory tidal volume (per breath)
VE TIDAL
mL
Expiratory tidal volume (per breath)
VE MIN
LPM
Expiratory minute volume (per minute)
PEEP H
cmH
2
O
High PEEP
PEEP L
cmH
2
O
Low PEEP
F PEAK
LPM
Maximum inspiratory flow
RESP.R
BPM
Respiratory
RR SPONT
BPM
Respiratory rate in spontaneous breath
VEMIN
SPONT
LPM
Expiratory volume per min in spontaneous breath
TI
sec
Inspiration time
TE
sec
Expiration time
I : E
Inspiration to expiration ratio
Ex.End Flow
LPM
Expiration flow
RSBi
b/min/mL
Rapid shallow breathing index
O
2
%
Oxygen concentration in volume %
CL
mL/ cmH
2
O Compliance
WOBv
J/L
Work of Breathing Ventilator
PR
BPM
Pulse Rate
SpO2
%
Oxygen saturation in blood