53
Cleaning and disinfection of the device between different patients is necessary and
should be done in accordance with the guidelines below:
1. Unplug the device from the external power adapter and external vacuum source.
2. Inspect the unit for any signs of contamination.
3. Gently wipe the LCD screen using a soft cloth and a designated LCD detergent. The
cloth should be damp but not dripping.
4. Using a soft cloth and 70% alcohol, gently wipe down the outer surface of the device.
5. Use a low-level cleaning agent or disinfectant and ensure its compatibility with
plastics. Follow manufacturer’s guidelines for use of cleaning agents.
6. Use another fiber-free damp cloth to remove any remaining solution from the device
surface.
7. Use a clean and dry soft cloth to wipe the outer device surface.
8. Do not use plastic solvents or abrasives.
Warning: Fluids should not be allowed to enter into the device. Do not use
unnecessary liquids to clean the device. If any liquids penetrate the device contact your
local authorized provider.
Maintenance Procedures
In order for the Vcare
®
unit proper and safe operation, routine maintenance and
system components inspection is required. The maintenance procedures should be
done only by trained and authorized technical personnel.
Prior to operation
Responsibility: operator.
1. Make sure that the external wall suction tube is not damaged and properly connected
to the unit.
2. Inspect the unit for any signs of contamination.
3. If there is any sign of contamination, clean the unit according to cleaning the unit
section in this manual.
Warning: do not connect the Vcare
®
unit to a damaged external power adapter.