31
For advanced Upper Limit settings, press FUNC shortly. This
allows the user to change the Upper Limit to a different value in
the pre-set available range.
The upper limit is displayed in [a].
Use the up and down scroll arrow buttons in the control panel to
change the displayed value and determine the required upper
limit.
Press OP/SET button to finalize your selection. The displayed
value stops flickering.
Press ↓ to continue to the next parameter.
Warning: do not set high values of upper limit in order for
the system to detect improper vacuum applied to the wound.
In case the vacuum applied by the device will be higher than
the upper limit value, an alert will be generated (see Alarms
chapter for more details).
5.1.3.2.4. Working Time Determination
When entering the Working Time determination function, the
following screen is displayed:
Insert printout
Common working time values are displayed in the first line: 1-5
minutes. The default working time value, 3 min, is flickering.
Use the up and down scroll arrow buttons in the control panel to
change the flickering value and determine the required working
time.
Press OP/SET button to finalize your selection. The value stops
flickering.
For advanced work time setting, press the FUNC button shortly.
The work time value will be displayed in [a].
Use the up and down scroll arrow buttons in the control panel to
change the displayed value and determine the required working