18
Primary Unit Functions Setting
Positioning of the Vcare
®
unit
The Vcare
®
unit should be placed close to the patient bed, in proximity to the external
vacuum and electricity sources. The unit should be preferably operated by direct
connection to a grounded electrical source and external vacuum source. The unit can be
placed on a stable flat dry surface or can be attached and secured to a designated
mobile stand
to enable patient ambulation.
The device must be operated in a quite environment with background noise of no more
than 45-50dB.
Make sure that the external power adapter and its connecting cables are not
creating a tripping hazard.
Make sure that the ventilation holes at the back of the unit are not covered or
blocked.
If the device is connected to mobile stand there is a need to make sure that it is
securely connected to the mobile stand and that the lock-pin is closed.
The Vcare
®
unit and tubing should be placed in a visible location in order to enable
direct eye contact with the system and enable early detection of bleeding and/ or
effective vacuum application to the treated area
.
Collection Canister installation
1. Insert the collection canister into the canister socket in the Vcare
®
unit by pressing
it inwards. When properly inserted, a clicking sound is heard. Make sure that the
canister is placed secured and sealed in the canister socket for effective operation of
the vacuum.
2. Connect the Proximal Canister tube to the Canister tubing outlet.
Replacing the Collection Canister
The Collection Canister may be replaced during normal operation, alarm condition or
routine maintenance.