31
time. The working time can be linearly adjusted in the range of 1-
15 min.
Press OP/SET button to finalize your selection. The displayed
value stops flickering.
Press ↓ to continue to the next parameter.
See the Clinical guidelines for recommended intermittent
treatment parameters.
For an intermittent therapy, the determined value is the
length of therapy's time segments.
For continuous therapy, choose the intermittent operation
mode and make sure that the device is connected to an
external vacuum source. This will enable the user to set
the pause time value to zero.
5.1.3.2.5. Pause Length Determination
When entering the Pause Length determination function, the
following screen is displayed:
Insert printout
Common pause time values are displayed in the first text row: 0-3
minutes. The default pause time value, 1 min, is flickering.
Use the up and down scroll arrow buttons in the control panel to
change the flickering value and determine the required pause
time.
Press OP/SET button to finalize your selection. The value stops
flickering.
For advanced pause time setting, press FUNC shortly. The pause
time value will be displayed in [a].
Use the up and down scroll arrow buttons in the control panel to
change the displayed value and determine the required pause
time. The pause time can be linearly adjusted within the range of
0-10 min.
Press OP/SET button to finalize your selection. The displayed
value stops flickering.